System Performance

ParameterSpecification
Sample Through-put 3.5 min/sample (6 hrs for complete 102 sample tray)*
Throughput 130-140 samples per 8-hour shift*
Solvent Usage 35 ml/sample**
Sample Batch Size 102 samples (tray has 104 positions, pos 1 = cleaning beaker, pos 2 = verification solvent)

* Sample through-put rate is directly dependent on the settings of the processing parameters. Specification stated is using default parameters.
** Solvent usage depends directly upon the dilution ratios and processing parameters. Specification stated is for 1:1 dilution using the default parameters.


Particle Sensor

ParameterSpecification
Model KLOTZ LDS 45/50 Laser Sensor
Measuring Range 4µm to 70µm (3µm to 200µm) oil calibration range
4µm co-incidence threshold 25,000 particles/ml (undiluted sample)
50,000 particles/ml (1:1 dilution)
Cell Dimensions 450µm x 500µm
Flow Rate 10 to 50 ml/min (CS-APC-2 instrument is calibrated at 30 ml/min)

Particle Counter

ParameterSpecification
Model KLOTZ USB Counter
Number of Channels 4,096

Sample Level Sensor

ParameterSpecification
Model Baumer UNKC 09
Accuracy ±0.1 mm from 3mm to 150mm

Physical Specifications

ParameterSpecification
Dimensions 36" (W) x 30" (H) x 31" (D) (91cm x 76cm x 79cm)
Weight 85 lbs (39 kg)
Voltage Requirement 100-120-230/240VAC selectable, 50/60 Hz.
Input Current 6.0A @100V, 5.0A @ 120V, 2.6A @ 230V, 2.5A @ 240V